IN-C005-1 hospital medical equipment cheap patient monitor for sale

  • FOB Price:US $300-500 / Set(s)
  • Min.Order:1 Set(s)
  • Production Capacity:1000sets one moneth
  • Payment Terms:T/T , PayPal , Small-amount payment , Money Gram , EXW , Others(alibaba pay)
  • Favorite
Icen Technology Company Limited

Business Type:Manufacturer

Country/Region:China

Ddu Verified

HOT Rank

8/10

Product Information

  • Model:IN-C005-1
  • Certification:CE
  • Net Weight:4KG
  • Dimension:42*32*43 cm
  • Packaging Details:6kg (with battery)
  • Brand Name:ICEN
  • Trademark:ICEN
  • Place of Origin:CHINA
  • Warranty:1 year

Description

IN-C005-1 hospital medical equipment cheap patient monitor 

 

 

Feature:

12.1'' screen with true color, wide viewing angle, high brightness LCD display.

HR/PR, NIBP, RR, ECG, SpO2, TEMP

Simple and friendly operating display interface.

Rechargeable lithium battery provides convenience for patients' moving.

Playback and browse function for long term waveform and monitor data record.

Optional printing output function, alarm triggers printing.

Auto double alarm with audible and visible signals

Anti-defibrillation, anti-interference from high frequency electric knife

A full-synchronistic lead multi-channel ECG display...

This medical equipment can be used in ICU, CCU and operation room.

Suitable for adult,pediatrics and neonate patients. 

Standard:

l HR/PR, NIBP, RR, ECG, SpO2, TEMP, Rechargeable lithium battery

Optional:

l Dual TEMP,l Dual IBP,l ETCO2,l Thermal recorder,l Wall mount,l Trolley

 

Detailed:

ECG

Lead configuration

Standard 3-lead or 5-lead 

3-leadRA,LA,LL,Lead method:I,II,III

5-leadRA,LA,LL,RL,V,Lead method:I,II,III,aVR,aVL,aVF,V

Increase ´250,´500,´1000,´2000

HR

Range

  Adult                   15 ~ 300bpm (beat/minute)

  Neonatal baby/children       15 ~ 350 bpm(beat/minute)

Precision  ±1% or ±1bpm,the larger prevails

Resolution   1 bpm(beat/minute)

Sensitivity   > 200 uV (Peak-to-peak value)

Input Impedance  > 5 (megohm)

Bandwidth

Diagnostic mode0.05~130Hz

Monitoring Mode    0.5~40Hz

Operation mode1~20Hz

Common Mode rejection Ratio

Diagnostic mode> 90 dB

Monitoring Mode    > 100 dB

Operation mode> 100 dB

Pole Polarization Vvoltage Range   ±300mV

Pacing Pulse Test

Test pacing pulse in accordance with the following conditions:

Amplitude:±2 mV ~ ±700mV

Width:    0.1ms ~ 2ms

Risetime:      10us~100µs

Pacing pulse inhibition

Amplitude:±2 mV ~ ±700mV

Width:   0.1ms ~ 2ms

Risetime:     10us ~ 100µs

Baseline Recovering Time  After defibrillation < 3 seconds 

Signal Range  ±8 mV (Peak-to-peak value)

Calibrating Signal  1mV(Peak-to-peak value),precision±5%

ST Segment Measuring Volume

Measuring range: -2.0mV ~ +2.0mV

No definition for other ranges.

RESP 

Measuring Method:RA-LL impedance

RESP Impedance Measuring Range:0.3~3Ω

Base Impedance Range:200~4000Ω

Bandwidth:0.1~2.5Hz

RESP Rate:

Range:

Adult          0~120BrPM

Children and neonatal baby  0~150 BrPM

Resolution    1 BrPM

Precision    ±2 BrPM

Asphyxia Alarm  10~40 seconds

 NIBP 

Measuring Method:Pulse wave oscillometry 

Work Mode:Manual/Automatic/STAT

Measuring Interval of Automatic Measuring Mode:

1,2,3,4,5,10,15,30,60,90,120,180,240,480 minute(s)

Measuring Time of STAT Mode:5 minutes

PR range:40 – 240 bpm

Overvoltage protection:

Adult mode:        300 mmHg

Children mode :       240 mmHg

Neonatal baby mode : 150 mmHg

SpO2 

Range:        0~100%

Resolution:  1%

Precision:   

70~100%:±2 DIGIT

0%~69%:no definition given

PR

Range:        20~300bpm

Resolution 1bpm   Precision :   ±3bpm










You Might Also Like
Change a group
Inquiry Cart(0)